Bellingham + Stanley Digital Refractometer RFM340-T with Peltier Temperature Control and Touchscreen

Share:


Bellingham + Stanley Digital Refractometer RFM340-T with Peltier Temperature Control and Touchscreen

Bellingham + Stanley Digital Refractometer RFM340-T with Peltier Temperature Control and Touchscreen

The RFM340-T combines high-precision peltier-temperature control and a dynamic touchscreen in one product, which is ideally suited for demanding applications in the food, beverage & chemical industries.
 

Advantages of the digital refractometer RFM340-T at a glance:

  • 7" high-resolution HD touchscreen display
  • Flat prism surface for easy cleaning
  • Wide beam scan for non-homogeneous samples
  • Brix precision to three decimal places* (6 d.p. RI)
  • US/EP/BP/JP compliant
  • Supports FDA Regulation 21 CFR Part 11
  • PHR-MEAN methods
  • USB & Ethernet connections

Bellingham + Stanley has been manufacturing refractometers for over a century. This vast experience has been incorporated into the development of the latest generation of instruments, resulting in the most advanced product range to date.

The RFM300-T is the latest in a series of refractometers designed specifically for use in harsh environments. With wide beam optics and one of the flattest prism platforms on the market, our RFM300-T refractometers are capable of measuring non-homogeneous samples such as fruit juices with pulp, chemical compounds and emulsions.

The ABS housing of the new device is ergonomically designed so that liquids spilled next to it can run off via the slopes or be collected by the PEEK drip skirt, which encloses the easy-to-clean stainless steel prism tray.

An inherent part of the modern design is the new high-resolution capacitive touchscreen, which not only allows use in factory environments and operation with gloves, but is also the centerpiece of the new software. This helps the user to quickly maneuver through the user and configuration menus.

SMART temperature stability, RFID user activation and application-specific, pre-programmed methods simplify routine analysis in the pharmaceutical and petrochemical industries. The included PC software allows the device to be controlled via an Ethernet connection, enabling remote technical support and training.

Specifications

Specifications RFM330-T RFM340-T
Scales
  Refractive Index
  Sugar (°Brix)
  User-defined 

1.32 – 1.58
0-100
100

1.32 – 1.58
0-100
100
Resolution
  Refractive Index
  Sugar (°Brix)

0.0001 / 0.00001 (selectable) 
0.1 / 0.01 (selectable)

0.000001 (selectable up to 6 d.p.)
0.01/0.001 (selectable up to 3 d.p.)
Accuracy
  Refractive Index

  
  Sugar (°Brix)

±0.00005


±0.04

±0.00002 (1.32 – 1.38 RI)
±0.00004 (1.38 – 1.58 RI)

±0.010 (0 – 30 °Brix)
±0.030 (30 – 100 °Brix)
Precision (Reproducibility)
Refractive Index
Sugar (°Brix)

± 0.00001
± 0.01

± 0.000005 (6 d.p.)
± 0.005 (3 d.p.)
User Scale Library 20+ pre-programmed scales including HFCS (3), sugar (4), sucrose SG (3), NaCl, Butyro, Wine Must (5), Urine SG (3), PHR-MEAN and more; plus customer programmable user scales via PC.
Reading Time Minimum 4 seconds (dependant on precision required)
Measuring Temperature Range 0°C or 10°C below ambient whichever is greater to 70°C
Temperature Sensor Accuracy ± 0.03 °C
Sample Temperature Stability ± 0.05 °C
Temperature Compensation
  Sucrose (°Brix)
  AG Fluids
  User-defined

5 – 80 °C
5 – 40 °C 
Simple coefficient (units/ °C) or polynomial function
Temperature Stability Checks None/delay time/repeatability/ Smart
(independently selectable by Method)
Sample Illumination Light Emitting Diode 589nm (100,000+ hours)
Prism / Dish Artificial Sapphire / Stainless Steel 316 / PEEK (spill barrier)
Prism Seal Silicon/Resin
Interfaces 3 x USB (A), 1 x Ethernet, RS232 via USB adaptor (optional accessory)
Power Instrument: 24 V DC, ±5%, <2A 
External PSU: 100-240V, 50-60Hz (supplied)
Humidity Range <90% RH (non-condensing)


 Inquiry - Bellingham + Stanley Digital Refractometer RFM340-T with Peltier Temperature Control and Touchscreen