Rocklabs Smart BOYD LSD

Share:


Rocklabs Smart BOYD LSD

Rocklabs Smart BOYD LSD

The Rocklabs Smart BOYD LSD combo combines the BOYD Crusher with a Linear Sample Divider (LSD) controlled by a Programmable Logic System (PLC) allowing an operator to pre-weigh samples and select the suitable split weights. It is capable of variable splits in one pass.

The Rocklabs Smart BOYD LSD achieves a higher degree of splitting accuracy and is able to split smaller samples with the same accuracy as larger samples. The operator can select two splits of different weights as well as a retained waste sample.

The combo can be equipped with a barcode scanner, have pre-set splitting programs and retain data/sample logging details.

 

Benefits:

  • No manual split adjustment
  • Less operator input
  • Optional waste feeder
  
 Specifications BOYD Crusher Linear Sample Divider (LSD)
 Motor power 5.5kW, 3 phase 0.09 kW, 3 Phase
 Maximum input lump size 70mm 10mm
 Product size -2mm (+80%) n/a
 Width x Height x Depth 694 x 1440 x 1070mm 380x816x792mm
 Weight 610kg 110kg
 Jaw width 300mm n/a
 Dust extraction 1 extraction port available (Ø50mm) 1 extraction port available (Ø50mm)
 Options
  • Raiser Platform & Stairs
  • Manual Sample Lifting Arm
  • Automated Dumpmaster Lifter & Large Capacity Hopper
  • Waste feeders or conveyor interface
  • Waste Sample Bagger
  • Data Logging Software
  • Keyboard & Barcode Scanner
  • Sample 1 Bin
  • Sample 2 Bin, or Bypass bin for single splitting
  • Waste Bin (or bypass bin for waste feeder/conveyor options)
  • Split configuration & weight or % selection are adjustable in software
 Warranty Rocklabs Smart BOYD LSD combo is sold with a 2000 operating hours or one year warranty (which ever comes first), covering all parts (excluding wear parts) and labour if used under normal laboratory conditions.  
 Wear Plates High Chrome Iron The wear plates are very resistant to abrasion. In many laboratories one set of plates will last for more than 1 year. Jaw plates can be turned top to bottom to extend the life.
     
    Standard deviation of samples %
 Dividing Cone and quartering 6.810
  Scoop sampling 5.140
  Riffle splitting 1.010
  Table sampling 2.090
  Rotating splitting 0.125
  Random variation 0.076